2.4.4 NDIR CO
Gas Detection
2
The key component for reliable TOC analysis is the precise and
interference-free detection of CO
very little maintenance because of its ability to handle the corrosive
gases generated in TOC analysis. This is due in part to its non-reflective
borosilicate sample cell. It has no moving parts which further reduces
the maintenance load and it uses an internal self-calibrating technique.
The NDIR unit is shown in Figure 2-10. Some of the features of the
NDIR unit are as follows:
• Specific, interference-free CO
• Dual-wavelength rationing compensates for drift
• Computer-controlled for accuracy
• Sapphire protected optics
• Non-corrosive, non-reflective sample cell (borosilicate)
• No moving parts or tools required for easy maintenance and
service
• No critical realignment required
The NDIR CO
2
wavelength ratioing technique with a single borosilicate sample cell that
Teledyne Analytical Instruments
. The Model 6750 NDIR unit requires
2

Figure 2-10: Model 6750 NDIR Unit

detector uses a completely solid state, dual
Theory of Operation
detection
2
17