Appendix 1. Additional Protective Grounding - FujiFilm CR-IR 357 Service Manual

Hide thumbs Also See for CR-IR 357:
Table of Contents

Advertisement

Appendix 1. ADDITIONAL PROTECTIVE
GROUNDING
Additional protective grounding will be required if installing this equipment in the patient
environment with the connected equipment in a non-medically used room. Refer to Medical
System Standard, IEC60601-1-1.
However, there may be cases where additional protective grounding is not possible.
(For example in cases like non-metallic faceplate screws as well as boxes and isolated
supply systems etc.) If this should happen, please refer to "Counter-measures for additional
protective grounding" on the next page.
This installation work should only be performed by persons certified to handle electrical
facility construction.
Definitions of "patient environment", "medically-used room", and "non-medically used room"
are given below.
Patient Environment
Patient environment is the area for the patient to receive medical procedures (treatment,
tests, diagnosis, monitoring). It is the space measuring 2.5 m in all four directions and 2.5 m
in height from the area of the patient's body.
It excludes the space traveled by the patient to reach the medically-used room.
Medically-Used Room
Room equipped with protective grounding (medical use outlet or medical use grounding
terminal) implemented by the medical grounding method. The protective grounding inside
the medically-used room is equipotential, and the protective grounding of this medically-used
room is equipotential to that in the other medically-used room.
Generally, a portion of the medically-used room is the patient environment.
Non-Medically Used Room
Areas outside the medically-used room are considered the non-medically used room.
014-204-01E
11.10.2005 FM4750
Connection Locations of Cables for Additional Protective Grounding of the
Equipment and Clamp Locations
INSTRUCTION
Please use power supply cord and clamp (which is required for keeping the additional
protective ground cord in position) approved by the electrical standard of the country to which
the equipment is to be installed.
<Counter-measures for additional protective grounding>
1. Engage the services of a professional to make the necessary preparations for the
connection of additional protective grounding.
2. If the connected device possess a hospital grade earth terminal and is mobile, move the
connected device and replace with a power supply plug which has hospital grade earth
terminal.
3. Install a isolating transformer to the connected device. In this way, the power supply from
the connected device will be separated from the main power supply.
CR-IR 357 Service Manual
Appx_IN-1