9.2
Image/Conveyance Checks
 NOTE 
Because the "MENU" and so forth differ depending on the IP type, image checks should be
performed with the "MENU" specified below.
1.
Read the IP prepared in "9.1 Check before procedures" by the CR Console menus
"QC/TEST" and "Image Format" or "Image Format-2" and "SINGLE", and output
an image.
014-204-01E
11.10.2005 FM4750
2.
Verify that there is no "unusual nonuniformity" found on the output film or on the
image displayed on the image monitor.
 CHECK 
There should be no uneven density difference in the main scan direction.
If there is any uneven density difference, compare the films outputted from the two IPs
of the same size, and determine whether such nonuniformity is caused by the machine
or X-ray tube.
 If there is uneven density difference in the same direction on two films
→ The machine may be the cause of such nonuniformity.
{Troubleshooting}
 If positions of uneven density difference are different
→ The X-ray tube may be the cause of such nonuniformity.