3.
Registering the Machine Information (CONNECTING
EQUIPMENT)
1.
Click on [CONNECTING EQUIPMENT] in the "Setup Configuration Item" window.
2.
Click on
.
→ The "Connecting Equipment" dialog box appears.
3.
Register the machine information as shown below.
I. Select "Reader".
II. Select the host name (e.g., ru0).
III. If several RU's are connected, select the second and subsequent host name(s) in
"Equipment #2" and so on.
4.
Click on
.
→ Registering the machine information is now completed.
The "Setup Configuration Item" window appears back on screen.
014-204-01E
11.10.2005 FM4750
4.
Saving Configuration and Exiting Service Utility
1.
From the "Config (F)" menu in the "Setup Configuration Item" window, select
"Save (V)".
→ The save confirmation window appears.
2.
Click on
.
→ The setup is saved.
3.
From the "Config (F)" menu, select "Close (C)".
→ The "Service Utility" window appears back on screen.
4.
Exit the Service Utility.