3.
Check the image format.
Make sure that the white blank portion is 2 mm or less, as actually measured on the IP,
from the outermost edge of the image frame of the film outputted.
CHECK
The white blank portion should be 2 mm or less.
If anything abnormal is found:
{Troubleshooting}
NOTE
The distance measured on the film may differ from the actual size on the IP, depending
on the reading size and film size. The actual size on the IP should be calculated using
the distance measured on the film and the reduction factor for the film. Compute the
actual size on the IP according to the following equation.
014-204-01E
11.10.2005 FM4750
CR-IR 357 Service Manual
IN-37
IN-37