When preparing, using, and disposing of chemicals, be sure to follow manufacturer recommendations and
local regulations.
WARNINGS
Follow the manufacturer's cleaning label instructions for solution strengths and
Cautions
Check the expiration date on all cleaning and disinfection solutions and wipes. Do
Do not allow cleaning solution or disinfectant into the battery compartment,
Do not use strong solvents, such as thinner or benzene, or abrasive cleansers since
Use only a soft cloth lightly dampened in a mild soap or compatible cleaning

Cleaning and disinfecting the system and protective case

The following procedures can be used to clean and disinfect the system and protective case.
Caution
To clean the system and protective case
1 Make sure the system is turned off.
2 Disconnect the transducer.
3 Remove the system from the protective case.
4 Using a soft, damp cloth (dampened with cleaner or disinfectant) or a premoistened disinfectant wipe,
clean any particulate matter or bodily fluids from all system and protective case surfaces.
Be sure to use only cleaners listed in
Wipe from clean areas to the soiled areas to avoid cross-contamination.
WARNINGS
Cleaning and disinfecting
Residual chemicals from some disinfectants can cause an allergic reaction in some
individuals.
disinfectant contact duration.
not use chemicals or wipes that have expired.
system controls, or transducer connector.
these will damage the exterior surfaces. Use only FUJIFILM SonoSite recommended
cleaners or disinfectants.
solution to remove any disinfectant on metal surfaces.
Do not immerse the system or the protective case in any liquid.
Table 2-2, "Compatible cleaners and
Wear the appropriate personal protective equipment (PPE), such as eyewear and
gloves, recommended by the chemical manufacturer.
disinfectants"on page 7.