Final Placement; Securing The Machine - FujiFilm CR-IR 357 Service Manual

Hide thumbs Also See for CR-IR 357:
Table of Contents

Advertisement

6.

Final Placement

For final placement of the machine, check the requests of the user, and observe the
procedures described below to secure the machine and cables in place.
WARNING
At an institution where a long cassette is used, be sure to fasten the machine to the
floor with the anti-topple retainer.
{IN: Appendix
2_Securing the Machine with the Anti-Topple Retainer}
 Retaining the machine with the adjustable feet;
When the machine is to be retained with the adjustable feet, refer to "6.1 Securing the
machine".
{IN:6.1_Securing the Machine}
 Retaining the machine by the anti-topple retainer;
When the machine is to be fixed by the anti-topple retainer (optional), refer to "Appendix 2.
Securing the Machine with the Anti-Topple Retainer".
{IN: Appendix
2_Securing the Machine with the Anti-Topple Retainer}
 Using the supporter
When using the supporter (optional), see "Appendix 3. Assembling and Securing the
Supporter."
{IN: Appendix
3_Assembling and Securing the Supporter}
014-204-01E
11.10.2005 FM4750
6.1

Securing the Machine

1.
Secure the four adjustable feet down onto the floor.
 NOTE 
When securing the adjustable feet, pay attention to the following points.
- The four adjustable feet should be evenly loaded.
However, if there is too little space on the right and left sides in the machine installation
place to secure the adjustable feet on the rear side, the two adjustable feet on the front
side should be secured in place. In this case, the two adjustable feet and two casters
should be evenly loaded.
- When the adjustable feet are secured, the casters should be detached from the floor.
2.
Secure the adjustable feet.
CR-IR 357 Service Manual