14.
Once pull out the machine. Loosen the button bolt, remove the spacer, and
fasten the button bolt to retain the bracket. After the bracket is retained, move
the machine, and again check as instructed in the procedure 13.
 NOTE 
When the spacer is removed, always take the procedure 13 to check the aperture.
014-204-01E
11.10.2005 FM4750
15.
While pushing the bracket retained on the floor against the machine, additionally
fasten the button bolt to retain the bracket mounted on the floor. When the
bracket retained on the floor does not move, once loosen the button bolt, push
the bracket against the machine, and fasten the button nut to retain the bracket
on the floor.