Setting up an 8 CC 28 GHz EVM Measurement

This procedure shows you how to USE the N9040B signal analysis 5G NR
X-Series application to observe EVM on an eight-carrier waveform.
Ensure the equipment and triggers are properly configured. Refer to
"Configuring the Equipment Setup" on page 7
the N9040B using 5G NR Mode" on page
Using the graphical user interface
On the VXG:
1. Select Preset to set the VXG to a known state.
2. Set Frequency to 28 GHz and Amplitude to 5 dBm.
3. Select the Signal block to open.
4. Set Vector Modulation Signal Mode to Waveform File.
5. Select the Markers block and set Sync to Marker 2 and then select Back to
64
close the Markers Setup.