5G NR Measurements Using the VSA Software
5G Waveform and EVM Analysis Using VSA Software
From the VSA menu panel, select File > Recall Recording and navigate to the
waveform file. Refer to the Keysight 89600 VSA Software Help for more
information.
2. From the menu bar, select Input > Channels > Baseband > 2 Channels.
3. In both the Ch1 and Ch2 Spectrum windows, set the Center Frequency to
M9383B/84B Measurement Guide
You will need to copy the 5GNR_UXR_Recording_28GHz.csv file to the
N9040B or the PC running the VSA application.
28 GHz, Span to 122.88 MHz, and then select Auto Range All
Auto Range samples the current input signal and then sets the full scale
input range to the minimum range that includes the peak voltage sample
of the input signal.
157