Basic Measurements
Making Measurements
Setting Up Frequency Modulation
Ensure the equipment and triggers are properly configured. Refer to
Equipment Setup" on page 7
page 9
Follow the steps below for FM analog modulation using the internal or an
external I/Q modulation source.
Using the graphical user interface
On the VXG:
1. Set Frequency to 20 GHz and Amplitude to 0 dBm.
2. Select the Signal block.
3. Select the Vector Modulation Signal Mode dropdown and select Analog
4. In the Analog Modulation Signal Setup:
M9383B/84B Measurement Guide
.
Modulation.
a. Set Modulation Type to FM.
b. Set Waveform to Sine.
c. Set Rate to 400 Hz.
d. Set Deviation to 10 MHz.
"Setting Up Triggers on the N9040B" on
and
"Configuring the