Keysight M9383B Measurement Manual page 32

Microwave signal generator
Hide thumbs Also See for M9383B:
Table of Contents

Advertisement

Basic Measurements
Making Measurements
Using the equivalent SCPI programming example
On the VXG:
SYSTem:PRESet
RF1:FREQuency:CW 20GHZ
RF1:POWer:AMPLitude 0dBm
SIGNal1:MODE WAVeform
SIGNal1:WAVeform:TRIGger:SYNC:MARKer M2
SIGN1:WAV " D:\Users\Instrument\Documents\Demo
Waveforms\GSM_1C_Burst.WFM"
SIGNal1 ON
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the N9040B:
FREQuency:CENTer 20GHZ
FREQuency:SPAN 900KHZ
ACPower:BANDwidth 470 Hz
DISPlay:TXPower:WINDow1:TRACe:MAXHold ON
On the VXG:
SIGN1:WAV " D:\Users\Instrument\Documents\Demo
Waveforms\WDCMA_TM1_64DPCH_4C.WFM"
SIGNal1 ON
RF1:OUTPut:STATe ON
On the N9040B:
ACPower:BANDwidth 620 Hz
FREQuency:SPAN 30MHZ
32