Keysight M9614A Startup Manual

Pxie 5-channel precision source/measure unit (smu)
Hide thumbs Also See for M9614A:

Advertisement

Quick Links

Keysight M9614A/M9615A
PXIe 5-Channel
Precision Source/Measure Unit
Startup Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight M9614A

  • Page 1 Keysight M9614A/M9615A PXIe 5-Channel Precision Source/Measure Unit Startup Guide...
  • Page 2 - This product complies with POLLUTION BUT NOT LIMITED TO THE IMPLIED WAR- other things, that Keysight: (1) Furnish techni- DEGREE 2 defined in IEC 61010-1. RANTIES OF MERCHANTABILITY AND FIT- cal information related to commercial com- NESS FOR A PARTICULAR PURPOSE.
  • Page 3 Return the instrument to a Keysight Technolo- Hot surface. Avoid contact. Sur- gies sales or service office for services and faces are hot and may cause per- repair to ensure that safety features are main- sonal injury if touched.
  • Page 4 The Chinese mark for paper- To return unwanted products, contact your documentation. The equipment based packaging materials; local Keysight office or visit the following will be marked with this symbol Paperboard and Corrugated website for more information. when it is necessary for the user Fiberboard http://about.keysight.com/en/companyinfo/...
  • Page 5: Table Of Contents

    Keysight M9614A/M9615A PXIe Precision Source/Measure Unit Startup Guide Contents Introduction Related Documentation Step 1: Unpack and Inspect the Module Electrostatic Discharge (ESD) Precautions Before Unpacking Inspect for Damage Return the Module for Service Step 2: Verify Shipment Contents Step 3: Install the Software...
  • Page 6 Guarding Installing Interlock Circuit Maintenance Cleaning Self Test Self Calibration Calibration...
  • Page 7: Introduction

    Keysight Connection Expert and a soft front panel (SFP). M9614A PXIe 5-Channel SMU, 500 kSa/s, 100 pA, 30 V, 500 mA • M9615A PXIe 5-Channel Precision SMU, 500 kSa/s, 10 pA, 30 V, 500 mA •...
  • Page 8: Step 1: Unpack And Inspect The Module

    Step 1: Unpack and Inspect the Module Step 1: Unpack and Inspect the Module Keysight’s PXIe chassis and instrument modules are shipped in materials which prevent static electricity damage. The modules should only be removed from the packaging in an anti-static area ensuring that correct anti-static precautions are taken.
  • Page 9: Before Unpacking

    Step 1: Unpack and Inspect the Module Before Unpacking Keysight M9614A/M9615A module is designed to meet IEC/EN61010-1 and must only be installed in the PXIe chassis which is certified by a Nationally Recognized Testing Laboratory. The PXIe chassis must already be installed on a static-safe workbench so that you can install the module soon after unpacking.
  • Page 10: Return The Module For Service

    Should it become necessary to return a module for repair or service, follow the steps below. 1. Review the warranty information shipped with your product. 2. Contact Keysight to obtain a Return Material Authorization (RMA) and return address. For assistance in finding Keysight contact information, visit www.keysight.com/find/assist.
  • Page 11: Step 2: Verify Shipment Contents

    PXIe 5-Channel Precision Source/Measure Unit or M9615A Quick Startup Poster Certification of calibration (without test data) Short bar Connector-terminal block 2.5 mm 5-terminal Consumable supplies of the M9614A/M9615A are shown in below. Table 1-2 Consumable Supplies Part number Description Quantity...
  • Page 12: Step 3: Install The Software

    M9614A/M9615A software. 1. Copy the M9614A/M9615A software installer into the controller. 2. Launch the M9614A/M9615A software installer. 3. Follow the installer prompts to install the M9614A/9615A software. 4. Reboot the controller when prompted by the M9614A/M9615A software installer.
  • Page 13: Step 4: Install The Module

    Step 4: Install the Module Current capacity of the PXIe chassis may restrict the module configuration. For more information, refer to the M9614A/M9615A data sheet and the PXIe chassis documentation. PXIe hardware does not support “hot-swap” (changing modules while power is applied to the chassis) capabilities.
  • Page 14 For more information on the module installation, refer to the PXIe chassis documentation. Ensure that filler panels are installed in all empty slots. Missing filler panels will impact cooling of the chassis and may cause RFI (radio frequency interference) with other devices. Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 15: Step 5: Verify Operation Of The Module

    Device Manager. The module should be visible in the device tree. The controller might request a reboot. Reboot the controller, if requested. 6. Check if the module is also visible in the Keysight Connection Expert; e.g. via Start menu > Keysight Connection Expert.
  • Page 16: To Update The Firmware

    If a firmware update is available, perform the following procedure to update the firmware of the M9614A/M9615A modules. 1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The Connect to Instrument dialog box will open. 2. In the list on the window, highlight the M9614A/M9615A modules to connect, and click Connect to launch the Keysight M960x PXIe Source Measure Unit Soft Front Panel (the M960x SFP).
  • Page 17: To Perform Self Test

    Before performing the self test, turn the instrument output off and disconnect cables from the measurement terminals. 1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The Connect to Instrument dialog box will open. 2. In the list on the window, highlight the M9614A/M9615A modules to connect, and click Connect to launch the M960x SFP.
  • Page 18: Front Panel Features

    Front Panel Features Front Panel Features This section describes the connectors and the indicators on the M9614A/M9615A front panel. “LED Indicators” • “Measurement Terminals” • “Trigger and Interlock Terminals” • Figure 1-1 M9614A/M9615A Front Panel Status indicator Access indicator Trigger and Interlock...
  • Page 19: Led Indicators

    Front Panel Features LED Indicators The M9614A/M9615A module has two LED indicators, Access and Status. These indicators show the status of the module. See Table 1-3 for the meanings. If the emergency shutdown occurs, both indicators turn red. Table 1-3...
  • Page 20 Ch3, LF Low force for channel 3 Ch4, HF High force for channel 4 Ch4, LF Low force for channel 4 Ch5, HF High force for channel 5 Ch5, LF Low force for channel 5 Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 21 These terminals are connected together by using the short bar when the M9614A/M9615A is shipped from the factory. The short bar must be connected to perform the grounded measurement. If you want to perform the floating measurement, or if you want to connect another instrument’s ground to the Low force, remove the short bar, and leave the...
  • Page 22 For the floating measurement, do not apply voltage over 40 V to the Low force terminal. Failure to heed this caution may result in damage to the M9614A/M9615A. Do not apply current or voltage to the frame/chassis terminal. Doing so will damage the M9614A/M9615A. Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 23: Trigger And Interlock Terminals

    Trigger and Interlock Terminals The M9614A/M9615A has the following terminals for the trigger signal input/output and the interlock circuit status detection. The M9614A/M9615A has a 5-pin terminal block on the front panel. The pin assignment of the terminal block is shown in Figure 1-1.
  • Page 24: Connecting A Dut

    Connecting a DUT Connecting a DUT This section describes how to connect a device under test (DUT) to the M9614A/M9615A, and includes the following descriptions. “2-Wire Connection or 4-Wire Connection” • “Floating” • “Guarding” • To avoid touching the end of the extension cable or the terminal area with the DUT, cover over the conductors with insulator.
  • Page 25: 2-Wire Connection Or 4-Wire Connection

    Using the interlock circuit The M9614A/M9615A provides an interlock function. If the interlock terminals are open, the M9614A/M9615A cannot apply a high voltage over the limit which is programmable, within42 V. The interlock circuit is not necessary for the M9614A/M9615A because the maximum output voltage is 30 V, but can be used to limit the output voltage to...
  • Page 26 Short bar High force High sense Low sense Low force Frame/chassis Figure 1-3 2-Wire Connection and 4-Wire Connection High force High force High sense Low sense Low force Low force 2-wire connection 4-wire connection Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 27: Floating

    Connecting a DUT Floating When the M9614A/M9615A is shipped from the factory, the Low force terminal is connected to the frame/chassis terminal by using the short bar. The short bar must be connected to perform the grounded measurement. If you want to perform the floating measurement or if you want to connect another instrument’s ground to the Low force, remove the short bar, and leave the...
  • Page 28 Do not connect the outer shield to anything other than the low terminal of DUT. Never connect the guard shield to any output, including the frame/chassis ground or any other guard terminal. Doing so will damage the M9614A/M9615A. Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 29: Installing Interlock Circuit

    42 V. The interlock circuit is not necessary for the M9614A/M9615A because the maximum output voltage is 30 V, but can be used to limit the output voltage to the specified value. If you set the threshold level to 0 V, the interlock function will be same as inhibit control.
  • Page 30 Mechanical switch (Keysight N1254A-402 or equivalent), 2 ea. • Connector-terminal block 2.5 mm 5-terminal, 1 ea. • BNC to ferrule terminal cable (Keysight PX0101A-001 (1.5 m), PX0101A-002 • (3.0 m), or equivalent), 1 ea. BNC connector (jack to soldering terminals), 1 ea.
  • Page 31 Installing Interlock Circuit Figure 1-6 Dimension of the Interlock Switch (Keysight N1254A-402) 10.3 4.75 35.6 Switch off 15.3 10.3 15.9 18.8 C OM 22.2 27.8 Units: mm Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 32: Maintenance

    Maintenance Maintenance This section describes the maintenance tasks of the M9614A/M9615A. “Cleaning” • “Self Test” • “Self Calibration” • “Calibration” • Cleaning To prevent electrical shock, disconnect the chassis from the mains before cleaning. Use a dry soft cloth or a soft cloth slightly dampened with water or a mild soap and water solution to clean the external surfaces of the chassis, modules, and accessories.
  • Page 33: Self Test

    Before performing the self calibration, turn the instrument output off and disconnect cables from the measurement terminals. To perform 1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The self calibration Connect to Instrument dialog box will open.
  • Page 34: Calibration

    It is recommended to perform the calibration once a year at least. For the calibration and adjustments, contact Keysight Technologies. Trained service personnel will perform the calibration and adjustments. Keysight M9614A/M9615A Startup Guide, Edition 1...
  • Page 36 This information is subject to change without notice. © Keysight Technologies 2020 Edition 1, May 2020 *M9615-90000* M9615-90000 www.keysight.com...

This manual is also suitable for:

M9615a

Table of Contents