5G New Radio (NR) Measurements using X-Apps
5G Waveform, EVM, and ACP Analysis Using X-Applications
To View the Results of Two Carriers
This example will show you have to view IQ Meas Time, Spectrum, and Error
Summary for carriers 0 and 1.
1. In the top right window, select the Window Title in the upper left hand
M9383B/84B Measurement Guide
corner of the window.
a. Select the Data dropdown and choose Demod > IQ Meas Time.
b. Select the Data dropdown again and select Component Carrier >
CC1.