5G New Radio (NR) Measurements using X-Apps
5G Waveform, EVM, and ACP Analysis Using X-Applications
9. For two channel instruments only: In the top right corner of the display,
The state of RF Out (All) does not change the RF Out state of a specific
channel. For example, if RF Out for Channel 1 is set to on and RF Out (All)
is set to off, no power is emitted from the RF Output connector of Channel
1. Likewise, if RF Out for Channel 1 is set to off and RF Out (All) is set to on,
no power is emitted from the RF Output connector of Channel 1.
On the N9040B:
1. Select Mode Preset to set 5 GNR mode to a known state.
2. Select Mode/Meas > 5G NR Mode > ACP Measurement.
If accessing the instrument via a Remote Desktop connection, select the
Screen tab (at the top of the display) to open the
Mode/Measurement/View Selector window.
3. Select Frequency and set Carrier Reference Frequency to 3.5 GHz.
4. Select Meas Setup > Advanced and turn Noise Correction On.
80
set RF Out (All) to On by selecting the switch.