5G NR Measurements Using the VSA Software
5G Waveform and EVM Analysis Using VSA Software
Using the equivalent SCPI programming example
On the VXG:
SYSTem:PRESet
RF1:FREQuency:CW 28GHZ
RF1:POWer:AMPLitude 5dBm
SIGNal1:MODE WAVeform
SIGNal1:WAVeform:TRIGger:SYNC:MARKer M2
SIGNal1:WAVeform "D:\Users\Instrument\Documents\MCS demo
waveforms\5GNR_8CC_FR2_120kHz_SCS_100MHz_Corrected_28GHz.wfm
"
SIGNal1 ON
DM:OPTimization:CHANnel EVM
RFALl:OUTPut ON
RF1:OUTPut ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the N9040B
INSTrument:SELect VSA89601
MMEMory:LOAD:DEMO
"D:\Users\Instrument\Documents\NR5G\data\NR5GEvm\CarrierSetu
p\1CC_FR2_120kHz SCS_100MHz_256QAM_DC Punctured_28GHz.setx"
INITiate:PAUSe
INPut:TRIGger:STYLe "External"
INPut:TRIGger:LEVel:EXTernal 1V
INPut:EXTension:PARameters:SET "ExtTriggerLoc", 2
INPut:EXTension:PARameters:SET "PhaseNoiseOptDualLoop", 1
NR5G:RLENgth 10
NR5G:SUBFrame:INTerval 2
NR5G:FRAMe:TRIGger:ENABled 1
NR5G:CAGGregation:CONFigure "Contiguous8CC"
NR5G:DC:PUNCtured 1
NR5G:MCFilter:ENABled 1
114