Keysight M9383B Measurement Manual page 18

Microwave signal generator
Hide thumbs Also See for M9383B:
Table of Contents

Advertisement

Basic Measurements
Making Measurements
Using the equivalent SCPI programming example
On The VXG:
SYSTem:PRESet
RF1:FREQuency:CW 20GHZ
RF1:POWer:AMPLitude 0dBm
SIGNal:MODE:AMODulation
SIGNal1:AMODulation:TYPE AM
SIGNal1:AM:SHAPe SINE
SIGNal1:AM:FREQuency 100KHZ
SIGNal1:AM 50
SIGNal1 ON
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the N9040B:
FREQuency:CENTer 20GHZ
FREQuency:SPAN 500KHZ
ACPower:BANDwidth 100 Hz
CALCulate:MARKer1:MODE DELTa
CALCulate:MARKer1:MAXimum:RIGHt
Repeat the above command until the marker is at the next highest peak.
To retrieve the delta marker:
CALCulate:MARKer1:Y?
18
M9383B/84B Measurement Guide