Keysight M9383B Measurement Manual page 21

Microwave signal generator
Hide thumbs Also See for M9383B:
Table of Contents

Advertisement

Basic Measurements
Making Measurements
Using the equivalent SCPI programming example
On the VXG:
SYSTem:PRESet
RF1:FREQuency:CW 20GHZ
RF1:POWer:AMPLitude 0dBm
SIGNal:MODE:AMODulation
SIGNal1:AMODulation:TYPE FM
SIGNal1:FM:SHAPe SINE
SIGNal1:FM:FREQuency 400KHZ
SIGNal1:FM 10
SIGNal1 ON
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the N9040B:
FREQuency:CENTer 20GHZ
FREQuency:SPAN 50MHZ
ACPower:BANDwidth 240 Hz
M9383B/84B Measurement Guide
21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M9384b

Table of Contents