Keysight M9370A Startup Manual

Pxie vector network analyzer modules

Advertisement

Quick Links

Keysight M937xA PXIe
Vector Network Analyzer
Modules
M9370A, 300 kHz to 4 GHz
M9371A, 300 kHz to 6.5 GHz
M9372A, 300 kHz to 9 GHz
M9373A, 300 kHz to 14 GHz
M9374A, 300 kHz to 20 GHz
M9375A, 300 kHz to 26.5 GHz
Startup Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight M9370A

  • Page 1 Keysight M937xA PXIe Vector Network Analyzer Modules M9370A, 300 kHz to 4 GHz M9371A, 300 kHz to 6.5 GHz M9372A, 300 kHz to 9 GHz M9373A, 300 kHz to 14 GHz M9374A, 300 kHz to 20 GHz M9375A, 300 kHz to 26.5 GHz...
  • Page 2: Safety Notices

    DOCUMENT OR ANY INFORMATION reproduce, release, perform, CONTAINED HEREIN. SHOULD display, or disclose commercial KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT computer software or WITH WARRANTY TERMS commercial computer software COVERING THE MATERIAL IN THIS...
  • Page 3: Table Of Contents

    3 ..........Safety and Regulatory Information Maintenance Shipment for Service Safety Symbols General Safety Considerations Before Applying Power Servicing Operating Conditions Keysight M937xA Startup Guide...
  • Page 4 Contents Regulatory Information Instructions for Use Safety Instrument Markings Declaration of Conformity Warranty Keysight M937xA Startup Guide...
  • Page 5: Getting Started

    Introduction The scope of this Startup Guide is to detail the processes of receiving and installing the network analyzer module and cables that compose the Keysight M937xA PXIe Vector Network Analyzer module. Additionally, installing the required software and verifying network analyzer module operation is documented.
  • Page 6: Inspect For Damage

    After unpacking a network analyzer module, inspect it for any shipping damage. Report any damage to the shipping agent immediately, as such damage is not covered by the warranty. To avoid damage when handling the network analyzer modules, do not touch exposed connector pins. Keysight Technologies Startup Guide...
  • Page 7: How To Return An Instrument For Service

    1. Review the warranty information shipped with your product. 2. Contact Keysight to obtain a Return Material Authorization (RMA) and return address. For assistance finding Keysight contact information, go to www.keysight.com/find/assist (worldwide contact information for repair and service).
  • Page 8: Step 3. Install The Software

    — Runs one of the operating systems listed in System Requirements (above). Remote controller (For Keysight M9018A chassis use only) A PC running one of the operating systems listed in System Requirements and a Keysight M9021A Cable Interface x8 with one of the following PC interface options: —...
  • Page 9: Software Installation Overview

    If you are using a remote controller, complete the following steps: 1. Install the cable interface between the remote controller and the chassis. If you are using an Keysight M9021A cable interface, refer to the Keysight M9021A documentation for further details.
  • Page 10 Getting Started STEP 3. Install the Software Software Installation Procedure 1. Install the Keysight IO Libraries Suite by downloading it from www.keysight.com/find/iosuite. Follow the installer prompts to install the IO libraries. 2. Install the network analyzer module software by downloading it from http://www.keysight.com/techicalsupport/main.
  • Page 11: Step 4. Install The Network Analyzer Modules

    Operating Conditions — Ensure proper chassis air flow is maintained. — The use of a Keysight M9018A chassis is recommended for optimum temperature control. If a different chassis is used, make sure it provides thermal protection if fans become inoperable or forced air cooling is obstructed.
  • Page 12: Tools Required For The Installation Procedure

    6. Holding the module by the injector/ejector handle, slide it into a slot, as shown in callouts 4 – 6 in Figure 1-3 on page a. Install the module into the slot of the chassis by placing the module card edges into the front module guides (top and bottom). Keysight Technologies Startup Guide...
  • Page 13 9. Install EMC filler panels and slot blockers into empty module slots to assure proper air circulation and operating temperatures. 10.If using only one M937xA module in the chassis, leave the 50 ohm loads attached as shown in Figure 1-4 on page Keysight Technologies Startup Guide...
  • Page 14 Leave the load installed on the LO Out port. c. On each of the remaining modules, remove both 50 ohm loads. d. Save the 50 ohm loads that were just removed for possible reuse in another module configuration. Keysight Technologies Startup Guide...
  • Page 15 16, between the Trig Out port and the Trig In port on each of the module pairs (1 & 2, 2 & 3, 3 & 4, 4 & 5, 5 & 6, 6 & 7, and 7 & 8). Keysight Technologies Startup Guide...
  • Page 16 STEP 4. Install the Network Analyzer Modules Figure 1-7 Installing Semi-Rigid Cable M9370-20018 Use the supplied cable removal tool (5002-3361) to remove the M9370-20018 cables if necessary. See Figure 1-8 on page Figure 1-8 Cable Removal Tool (part number 5002-3361) Keysight Technologies Startup Guide...
  • Page 17 Ref Out port and the Ref In port on each of the module pairs (1 & 2, 2 & 3, 3 & 4, 4 & 5, 5 & 6, 6 & 7, and 7 & 8). Keysight Technologies Startup Guide...
  • Page 18 STEP 4. Install the Network Analyzer Modules Figure 1-10 Installing Semi-Rigid Cable M9370-20019 Use the supplied cable removal tool (5002-3361) to remove the M9370-20019 cables if necessary. See Figure 1-8 on page You have now completed the hardware installation procedure. Keysight Technologies Startup Guide...
  • Page 19: Front Panel Features

    SMB female connector. Inputs a signal to trigger the network analyzer module. Trig Out SMB female connector. Outputs a signal either before or after a measurement. Trig Ready SMB female connector. Outputs a “READY’ signal to other devices. Keysight Technologies Startup Guide...
  • Page 20: Step 5. Verify Operation Of The Network Analyzer Modules

    6. If shorts and opens are not connected to all ports, you will be prompted to connect them as they are needed. 7. The result of the operator’s check will be shown as a PASS or FAIL next to each test (refer to Figure 1-11 on page Keysight Technologies Startup Guide...
  • Page 21: If The Operator's Check Fails

    1. Clean the test ports, and the open(s) or short(s). Torque to specification. Repeat the check. 2. If the check still fails, return the network analyzer module to Keysight. See “How to Return an Instrument for Service” on page Keysight Technologies Startup Guide...
  • Page 22 Getting Started STEP 5. Verify Operation of the Network Analyzer Modules Keysight Technologies Startup Guide...
  • Page 23: Getting Help With Your Network Analyzer Module

    The M937xA PXI Network Analyzer System Web page is another resource for help with the analyzer module. You can find all M937xA PXI network analyzer module documentation and other resources at: http://na.support.keysight.com/pxi/help. Check for software updates at: http://www.keysight.com/techicalsupport/main. Enter the model for your instrument and click Find...
  • Page 24: Contacting Keysight

    Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at: www.keysight.com/find/assist. If you do not have access to the Internet, please contact your Keysight field engineer. In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number.
  • Page 25: Safety And Regulatory Information

    Do not attempt to clean internally. Shipment for Service Contact Keysight Technologies for instructions on where to ship the analyzer module for service. Refer to “Contacting Keysight” on page Ship the analyzer module using the original packaging materials. Shipping the analyzer module in anything other than the original packaging may result in non-warranted damage.
  • Page 26: Safety Symbols

    Denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. Keysight Technologies Startup Guide...
  • Page 27: General Safety Considerations

    If modules are allowed to exceed the specified full temperature range, module surface temperatures can cause discomfort or burns if touched. If a module exceeds the full temperature range, always allow it to cool before touching it or removing it from the chassis. Keysight Technologies Startup Guide...
  • Page 28: Regulatory Information

    기로서 판 매자 또는 사용자는 이 점을 주 의하시기 바라 며 , 가정외의 지역 에서 사용하는 것을 목적으 로 합니 Safety Acoustic statement (European Machinery Directive) — Acoustic noise emission — LpA <70 dB — Operator position — Normal operation mode per ISO 7779 Keysight Technologies Startup Guide...
  • Page 29: Instrument Markings

    South Korean Certification (KC) mark; includes the marking's identifier code which follows the format: MSIP-REM-YYY-ZZZZZZZZZZZZZZ or KCC-REM-YYY-ZZZZZZZZZZZZ. Declaration of Conformity A declaration of conformity is available upon request, or a copy is available on the Keysight Technologies Web site at Keysight Technologies Startup Guide...
  • Page 30: Warranty

    The actual warranty on your instrument depends on the date it was ordered as well as whether or not any warranty options were purchased at that time. To determine the exact warranty on your instrument, contact Keysight Technologies with the model and serial number of your instrument. Refer to “Contacting Keysight”...
  • Page 31 Installation Note Xxxxx-xxxxx...
  • Page 32 This information is subject to change without notice. © Keysight Technologies 2014-2019 Edition 2, June 2019 www.keysight.com...