5G New Radio (NR) Measurements using X-Apps
5G Waveform, EVM, and ACP Analysis Using X-Applications
11.For two channel instruments only: In the top right corner of the display,
The state of RF Out (All) does not change the RF Out state of a specific
channel. For example, if RF Out for Channel 1 is set to on and RF Out (All)
is set to off, no power is emitted from the RF Output connector of Channel
1. Likewise, if RF Out for Channel 1 is set to off and RF Out (All) is set to on,
no power is emitted from the RF Output connector of Channel 1.
On the N9040B:
All example waveforms and setup files are located on the VXG at:
D:\Users\Instrument\Documents\Keysight\PathWave
\SignalGenerator\Examples
You will need to copy over the setup files to the N9040B or the PC running
the VSA application.
1. Select Recall (If accessing the N9040B remotely, select the Folder icon at
M9383B/84B Measurement Guide
set RF Out (All) to On by selecting the switch.
the bottom of the display) > Component Carrier Setup > Recall From > 
5GNR_1CC_FR2_120kHz_SCS_100MHz_256QAM_DCpunctured_
28GHz.scp 
Recall