5G NR Measurements Using the VSA Software
5G Waveform and EVM Analysis Using VSA Software
Using the equivalent SCPI programming example
On the VXG:
SYSTem:PRESet
RF1:FREQuency:CW 28GHZ
RF1:POWer:AMPLitude 5dBm
SIGNal1:MODE WAVeform
SIGNal1:WAVeform:TRIGger:SYNC:MARKer M2
SIGNal1:WAVeform "D:\Users\Instrument\Documents\MCS demo
waveforms
5GNR_1CC_FR2_120kHz_SCS_100MHz_256QAM_DCPunctured_28GHz.wfm"
SIGNal1 ON
DM:OPTimization:CHANnel EVM
RFALl:OUTPut ON
RF1:OUTPut ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the N9040B:
INSTrument:SELect VSA89601
All example waveforms and setup files are located on the VXG at:
D:\Users\Instrument\Documents\Keysight\PathWave
\SignalGenerator\Examples
You will need to copy over the setup files to the N9040B or the PC running
the VSA application.
SYSTem:PRESet
MMEMory:LOAD
"D:Users\Instrument\MyDocuments\NR5G\data\NR5GEvm\Carrier
Setup\5GNR_1CC_FR2_120kHz
SCS_100MHz_256QAM_DCPunctured_28GHz.setx"
INITiate:PAUSe
INPut:TRIGger:STYle "EXTERNAL"
INPut:TRIGger:LEVel:EXTernal 1V
INPut:EXTension:PARameters:SET "ExtTriggerLoc", 2
INPut:EXTension:PARameters:SET "PhaseNoiseOptDualLoop", 1
M9383B/84B Measurement Guide