Functional Checks (See Also Chapter 4) - GE LOGIQ 9 Technical Manual

Hide thumbs Also See for LOGIQ 9:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2294854-100, R
IRECTION
10-5-2

Functional Checks (See Also Chapter 4)

The functional checks take about 60 minutes to perform. Refer to the system user documentation
whenever necessary.
10-5-2-1
System Checks
Table 10-7
Chec
k
NOTE:
* Some software may be considered standard depending upon system model configuration.
10-8
PROPRIETARY TO GE
3
EVISION
System Functional Checks
Step
Verify basic B-Mode (2D) operation. Check the basic system controls that affect this mode
B-Mode
of operation.
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic system controls that
CF-Mode
affect this mode of operation.
Verify basic Doppler operation (PW and CW if available). Check the basic system controls
Doppler Modes
that affect this mode of operation.
Verify basic M-Mode operation. Check the basic system controls that affect this mode of
M-Mode
operation.
Verify the basic operation of all optional modes such as Multi-Image, 3D, Contrast,
*Applicable Software
Harmonics, Cine, Stress Echo,... etc. Check the basic system controls that affect each
Options
options operation.
Use the Visual Channel Utilityon the 10L probe to verify that all system xmit/recv channels
Xmit/Recv Elements
are functional.
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
Monitor
Verify basic Monitor display functions. Refer to Chapter 3 of the User Manual.
Touch Panel
Verify basic Touch Panel display functions. Refer to Chapter 3 of the User Manual.
Scan a gray scale phantom and use the measurement controls to verify distance and area
Measurements
calculation accuracy. Refer to the User Manual, Chapter 18, for measurement accuracy
specifications.
Section 10-5 - System Periodic Maintenance
LOGIQ™ 9 P
ROPRIETARY
Description
M
ANUAL