23. Attach the control knob
as follows:
– Slide the control knob
as far as it will go.
– Turn the knob until the white line points to
the lowest value.
– Hold the knob steady with the special tool
and screw it down.
24. Check the display on the control knob
when the knob is turned to the left stop the
white line must point to the lowest MV value,
when it is turned to the right stop the line must
point to the highest value. If this is not the case,
loosen the nuts and align the control knob
25.
Perform calibration (see "6.9 Calibration after re-
moval of PCB, ventilation control knob 5 and/or
pneumatic block" on page 25)
26. Close the device (see "6.5 Closing the
device" on page 23).
27. Turn the MEDUMAT Basic / Basic p round.
28. Place the lid
on the knob
34
29. Perform a functional check (see "3. Final
Check" on page 8).

6.16 Replace 3/2 solenoid valve

Tools required:
Crosstip screwdriver, size 2,
10 mm socket wrench,
Special tool WM 22829 from set WM 15349
If necessary, side nippers
If necessary, cable tie
Special pliers WM 22928
Calibration device WM 22836
Crosstip screwdriver, size 0
1. Open the device (see "6.4 Opening the
device" on page 22).
2. Remove the pneumatic block (see "6.15 Re-
placing the pneumatic block with
angled outlet" on page 34).
. To do so, proceed
35
onto the spindle
35
:
35
.
35
.
.
35
35
5
34
(Two-way switch only on Basic p)