Expiratory Flow Sensors Calibration - Siare Falco 202 User Manual

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

4.13.4 Expiratory flow sensors calibration

The Expiratory flow sensors calibration is necessary in the following cases.
Noted differences are more than 15% ( over 100ml ) between the set Volume value (
VTi - Vte ) and the expired Tidal Volume reading ( Vte ).
In case of first calibration of ventilator or replacement of flow sensors (INSP inside the
unit or EXP outside the unit), it is suggested to perform this Calibration.
Turn the encoder knob to select the parameter to
be calibrated: the activated message is
highlighted.
Press the encoder knob: the Expiratory flow
sensors calibration program is showed
Press START to begin the expiratory flow sensors calibration.
Press ESC to verify the expiratory flow sensors calibration.
4-58
User manual, DU3222102

Advertisement

Table of Contents
loading

Table of Contents