Siare Falco 202 User Manual page 184

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

Frequency
Every day / when
necessary
Every 2 weeks (at
least)
Every 6 months or
1000 working hours
( * )
Every year ( * )
Falco 202
Component
Oxygen sensor
Condensation trap
filter
Lung ventilator
Turbine air filter
EXP. V. Monoblock
Turbine air filter
Lung ventilator
Oxygen sensor
O
and Turbine air
2
filter
Patient circuit (silicone
tubes)
Washers / O-Rings
Lung ventilator
Procedure / Action
Calibrate according to the procedures described in
this manual.
Check for any water collection, drain and clean
when necessary.
General cleaning and checks.
Sterilize / disinfect according to the procedure
described in this manual and according to local
standards.
Replace. Components that cannot be destroyed
should be sterilized before disposal.
The lung ventilator must be inspected and
checked in general and any worn parts must be
replaced.
Use the appropriate preventive maintenance kit.
Replace. The working life of the cell depends on
the working environment. If the temperature or the
O
% is high, the working life of the sensor will be
2
lower.
Replace.
Sterilize according to the procedure described in
this manual and according to local standards.
Components that cannot be destroyed should be
sterilized before disposal.
Check the lung ventilator performance.
This includes an electrical safety test and
inspection of the ventilator for mechanical damage
and legibility of the labels.
The lung ventilator must also be inspected and
checked in general and worn parts must be
replaced, using the appropriate preventive
maintenance kit.
Use the appropriate preventive maintenance kit.
6-9