Press START to begin the expiratory flow
sensors calibration.
Press ESC to stop the expiratory flow
sensors calibration.
Falco 202
At the end of the Expiratory flow sensors calibration the system
come back to the " CALIBRATRION PROGRAMS " menu.
In case of first calibration of ventilator, we suggest to execute this
procedure after having checked the PEEP calibration and after
performing a Turbine characterization (see 4.13.3).
For more information please refer to the SERVICE manual.