3.6.5 Preliminary checks - Parameters monitoring

Falco 202
Based on the PRP set by the clinician and on the patient's [patient
simulator] characteristics, the lung ventilator is able to monitor and
measure a series of values necessary for the patient's clinical
evaluation.
Before checking the value of the set parameters, leave the ventilator on
for at least 15 minutes. This way the system will be able to reach its
operating condition.
5) Check the compliance
between the parameters set
and those monitored.
In the middle of the screen the
system displays the operating
curves.
On the top of the screen you
can see the monitored
parameters.
6) Change the PRP values.
PEEP : 5, 10 cmH
Tr. I : -1 cmH
O, 1 L/min
2
O
: 60%
2
7) Check the correspondence
between the monitored
parameters and the displayed
curves.
Note: Oxygen consumption
calculation
The oxygen consumption value is
showed in L/min for oxygen
settings of more than 21%.
O
2
3-27