Siare Falco 202 User Manual page 73

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

Conditions to be met for proper calibration:
the O
sensor must be placed in its seat
2
the O
must be electrically connected through the suitable cable
2
the medical gases must be properly connected.
If any of these conditions is not met, the calibration cannot be successful.
Falco 202
REPLACING THE OXYGEN CELL
The oxygen cell must be replaced when, at the end of the calibration
phase, appears a detected voltage value lower than 8mV and/or if the
system displays the relevant alarm message.
To order the replace sensor and to dispose of the worn one, please see
chapter 6 " Maintenance ".
WARNING !! Ventilator malfunctions risk
If the TEST result is negative check:
if the O
sensor is installed and electrically connected to the ventilator
2
if the O
sensor is worn out (the oxygen detection cell is worn out;
2
replace the O
sensor).
2
TESTS ON DEMAND FAILED
If the system does not exceed the preliminary checks phase, please see
chapter Alarms - Trouble Shooting or contact the nearest Siare Support
Centre or any other support centres authorised by Siare.
The oxygen sensor calibration
procedure was not completed
successfully.