4.13.3 Turbine characterization

The turbine characterization calibration is necessary in the following cases.
When you note differences out of tolerance (more than 2% of the end of the scale +
8% of the measured value) on the airways pressure values between the set and
measured values
Software update or first turning-on after an important repairing (for ex. turbine
replacement or inspiratory flow sensor replacement)
Turn the encoder knob to select the parameter to
be calibrated: the activated message is
highlighted.
Falco 202
CAUTION
In order to start the Turbine characterization calibration program is
necessary the intervention of qualified SIARE personnel or qualified
technical personnel authorized by SIARE.
For more information on the procedure please refer to the cfr. 3.8.1
and/or SERVICE manual.