Modification of the PRP
Turn the knob, select the parameter you want to change, press the
knob to activate the value editing function.
Turn the knob clockwise to increase (anticlockwise to decrease) the
value of the selected parameter.
Press the knob to confirm the modification.
For tests and checks, please use the patient simulator SIARE cod.
LS.AB.001 that is equipped with variable resistance and compliance.
3) Set the physiological
respiratory parameters (PRP)
Vti
500
RR
15
I:E
1:2
Tr. I
-1 cmH
O
2
1 L/min
0
Pause
OFF, 5, 10 cmH2O
PEEP
21%
O
2
4) Press the START button: the
ventilator begins its cycle.
User manual, DU322210210