Siare Falco 202 User Manual page 171

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

FiO
% low
2
Min. Vte / VM
Falco 202
This alarm is activated when the measured concentration of oxygen is
below the set limit.
Check that the oxygen cell is fitted correctly in its housing.
Check that the corresponding alarm limits are set correctly.
Calibrate the oxygen cell: if the problem occurs again after a short
time, replace the oxygen cell.
Check that the feeding pressure of the medical gases is correct: if
it is not, check the pressure of the distribution system and the
correct connection to the supply.
Check that the mask, endotrachéal tube and patient circuit are not
in some way clogged, bent or crushed. If this is the case,
eliminate the problem or replace them.
Contact the Siare Service Centre or a Centre authorised by Siare.
This alarm condition occurs in case the Vte is lower than set value.
Check that the corresponding alarm limits are set correctly.
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is
the case, eliminate the problem or replace them.
Check that the mask, endotracheal tube and patient circuit are not
in some way clogged, bent or crushed. If this is the case,
eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate,
I/E, Trigger).
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
Check if the lung ventilator works properly verifying the airways
pressure. If the lung ventilator works properly check the flow
sensor calibration and its connection with the device.
Contact the Siare Service Centre or a Centre authorised by Siare.
5-17