2.5.5 Graphic area

2-18
ALARM SILENCING
Press the ALARM RESET key to interrupt the acoustic signal for a pre-
set period of time.
During the silencing period, the text of the alarm will still be displayed.
Press the ALARM RESET key once again to delete the alarm text, only
if the alarm activation condition is no longer present.
If during the silencing period, a new alarm (of high priority) occurs, the
alarms silencing command is automatically cancelled and the acoustic
and visual signals are reactivated.
WARNING!
Patient injury hazard
Alarms silencing. The clinician should not interrupt patient control during
alarms silencing period.
The lung ventilator is equipped with tools for respiratory curves and loops
display so as to quickly and accurately notify the clinician on the patient's
condition.
User manual, DU3222102