Select the INSP HOLD mode and
press the encoder knob: the system
will extend the inspiration time from 5
up to 15 seconds.
The function activation is displayed
by the screen and signalled by the
yellow LED inside the box, that turns
on.
Select the EXP HOLD mode and
press the encoder knob: the system
will extend the expiration time from 5
up to 10 seconds.
The function activation is displayed
by the screen and signalled by the
yellow LED inside the box, that turns
on.
Falco 202
The clinician can enable both modes in sequence; the ventilator will carry
out both modes, giving priority to the mode that was activated first.
INSP HOLD and EXP HOLD modes:
are deactivated automatically
or by pressing the encoder knob.
By selecting the MAN mode using the encoder knob (at the bottom
of the display) the system provides the patient with a breath.
The breath ventilation parameters depend on the set operating
mode. The function activation is monitored by the screen and
signalled by the green LED inside the box, that turns on.
This mode is active while the lung ventilator is running.