LUNG VENTILATOR FUNCTIONAL FEATURES
Use destination
Operation principle
Pressure automatic
compensation
Dead space compensation
Automatic leaks
compensation
Leak % visualization
Respiratory parameters
default setting
Ventilation modalities
Breathing rate VC/VAC
Inspiratory Time / Expiratory
Time (maximum, minimum)
Breathing rate
V-SIMV e P-SIMV
SIMV Inspiratory time
Tidal volume
I:E ratio
Inspiratory pause
Inspiratory pressure limit
Falco 202
Falco 202 is a lung ventilator for use in home care and transport with patients
affected by respiratory diseases and it is suitable for ventilation of adult and
paediatric patients.
Time cycled at constant volume
Pressure cycled
Microprocessor controlled flow
Spontaneous breath with integrated valve
Automatic compensation of atmospheric pressure on measured pressure:
present
Automatic compensation of mechanical and patient circuit dead space
Max 60 l/min
Present
Present (Paediatric, Adult)
APCV, APCV-TV, PSV, PSV-TV (Auto Weaning), VC/VAC, V-SIMV+PS,
P-SIMV+PS, CPAP/PSV
SIGH, Apnea BACK-UP (NIV PSV, NIV PSV-TV, CPAP/PSV) , MANUAL
From 4 to 80 bpm
Ti min = 0.036sec (minimum inspiratory time)
Ti max = 9.6sec (maximum inspiratory time)
Te min = 0.08sec (minimum expiratory time)
Te max = 10.9sec (maximum expiratory time)
From 1 to 60 bpm
From 0.2 to 5.0 sec.
From 100 to 3000 ml (Adult)
From 50 to 400 ml (Paediatric)
From 1:10 to 4:1
From 0 to 60 % of the inspiratory time
Pinsp : from 2 to 80 cmH 2 O (in function of low and high pressure alarm set)
A-3