Low voltage power supply
A proper connector for low voltage power supply
line (12Vdc / 7A) is placed on the ventilator rear
side.
•
Insert the 12Vdc power supply cable plug to
the plug placed on the back of the ventilator.
Battery power supply
3-6
WARNING !! Personal injury - physical hazard
In order to avoid any electric shock hazard, make sure that the supply
cable is connected to an electrical socket with the grounding cable
connected.
• The lung ventilator complies with the requirements for electro-medical
devices detailed on chapter 1.5 ( Norms and standards regulations ).
• To ensure proper operation of the lung ventilator, please connect to it
only additional devices that comply with the standards specified above.
• The User should be always sure that the power supply electric plug is
accessible during the normal using and functioning of the device.
•
There should always be a battery installed inside the ventilator.
•
If there is no battery, the ventilator is not protected against voltage
drops or mains power supply cut off.
•
The ventilator should not be used without a charged battery.
•
The use with battery should be limited to short periods of time and
should not be considered an alternative to mains power supply.
•
Do not open the ventilator to replace the batteries or to carry out
maintenance operations on the same battery charger.
BATTERY LIFE
The battery operation time varies as follows: old battery or not fully
efficient, unusual ventilator parameters.
Replace the battery each two years (please refer to cap. 6 Maintenance).
User manual, DU322210210