Trends Storage capacity (72 h) of all measured parameters.
Events Memory storage up to 100 machine events including the alarms.
Patient data The patient data can be set and cancelled
Default parameters The default parameters can be restored
SETTING function (set of
physiological breathing
parameters)
Range of measured
parameters
Displayed parameters RR (bpm), I:E, ( * ) FiO 2 (%), Vte (ml), VM (L/min), PAW , PEEP,
Additional displayed
parameters
Displayed graphics
Flow sensor
Calibration Automatic (started by the operator)
Maintenance By steam or chemical disinfection
Oxymeter ( * )
Calibration Automatic (started by the Operator)
( * ) Optional : if FiO2 option is present
A-6
CPAP (cmH 2 O), Slope, I:E, RR(bpm), RRsimv (bpm), Pause (%), PEEP
(cmH 2 O), Pinsp (cmH 2 O), PMax - Pmin - PS (cmH 2 O), SIGH (% - bpm), Ti
(s), Ti Max (s), Tr. E (%), Tr. I (L/min - cmH 2 O ), Vte - Vti (ml), BACK-UP
parameters
•
Respiratory rate (range: 0 ÷ 200 bpm)
•
I:E ratio (range 1:99 ÷ 99:1)
•
( * ) % of O 2 (range: 0% ÷ 100%)
•
Tidal Volume: Vte, Vti (range: 0 ÷ 3000 ml)
•
Minute Volume (range: 0 ÷ 40 l/min)
•
PAW: peak, mean, plateau, PEEP (range -20 ÷ 80 cmH 2 O)
•
Inspiratory Peak Flow : Fi (range: 1 ÷ 190 l/min)
•
Expiratory Peak Flow : Fe (range: 1 ÷ 150 l/min )
•
Tinsp., Texp, Tpause (range 0.036 ÷ 10.9 sec)
•
Static compliance (range: 10 ÷ 150 ml/cmH 2 O)
CPAP (cmH 2 O)
MAP (cmH 2 O), Pplateau (cmH 2 O), Fi (L/min), Fe (L/min), Ti (sec.), Te (sec.),
Tpause (sec.), Cs (ml/cmH 2 O)
•
CURVES: Pressure - Flow - Volume
•
LOOPS : Pressure / Volume - Flow / Volume - Pressure/Flow
•
Auto range
Magnetic disturbance (patented), multi-usage type
Electronic (value displayed in breathing parameters)
User manual, DU3222102