Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

EXP V.
Monoblock
Couplings and
connectors
Turbine air
filter
6-6
WARNING !!
It is necessary to have at least one spare patient circuit in
stock for routine use and /or accidental breaks.
Disinfect with steam or chemically.
The EXP V. monoblock includes the expiratory valve and the
flow sensor.
WARNING !!
Do not attempt to dismantle or clean with compressed air.
The EXP. V. monoblock can be washed and disinfected by
immersing it in a bowl with 3 centimetres of liquid, keeping
the connector for the electrical connections facing upwards.
Dismantle and clean: sterilize in an
autoclave, disinfect with steam or
chemically.
Check that there are no splits and replace
them if they are damaged.
Dismantle and clean with hot water and a
neutral detergent solution.
Check that there are no splits and replace
them if they are damaged.
Do not clean or re-use disposable circuit tubes.
Do not clean or re-use if the filters are the disposable
type.
Components that cannot be destroyed should be sterilized
and disinfected according to local standards.
Risk of injury for the patient
It is possible to sterilize the
component with gamma rays
or ethylene oxide (ETO).
Risk of device failure.
Before using again, eliminate
any humidity inside the
components by means of
compressed air.
Before mounting the filter
again, eliminate any humidity
inside the components by
means of compressed air.