ALARMS
Alarm types
Alarm default setting
Alarm priority
Alarms with limits set up by the operator
Airways pressure
Respiratory rate High – Low
Expired tidal volume High – Low
Expired minute volume High – Low
FiO 2 concentration High – Low
Electric power supply Alarm occurs in case of failure of external power supply
Apnea Low Rate (function of Apnea BACK-UP)
System alarms
Level (charge)
Level (charge)
Battery level (low)
Disconnected battery
Battery over temperature
Circuit disconnected
Turbine fault
Turbine over temperature
Turbine over current
CAN BUS error
Maintenance
Falco 202
By MENU: with limits set by the operator
By DEFAULT: the operator cannot set them up
Present (Paediatric, Adult)
High - Mean - Standby
High – Low
PEEP High – Low
Battery at 50%
Battery at 25%
10 Minutes
Yes / No
Indication of exceeding the temperature limits inside the battery
Indication of patient circuit disconnected
Signals in case of a blower fault condition
Indication of exceeding the temperature limits inside the turbine
Indication of exceeding the current limits inside the turbine
Electronic boards CAN connection wrong
2000 hours
A-7