O
2
Concentration percentage delivered to the patient can be set
from 21% to 100% .
Pause (%)
Inspiratory pause time. The "inspiratory pause time" is displayed
on the screen in % (% of the inspiratory time). It is also used to
calculate the lung mechanics parameters (resistance and static
compliance).
PEEP (cmH
O)
2
Positive airway pressure value during expiratory phase.
Pinsp (cmH
O)
2
Maximum airway pressure limit value. The parameter is used in
pressure controlled modes to fix an operating limit for the airway
pressure that shall not be exceeded.
PMax (cmH
O)
2
Maximum airway pressure limit.
Pmin (cmH
O)
2
Minimum airway pressure limit.
Pressure Low (cmH
O) - Pressure High (cmH
O)
2
2
Pressure levels to be set in APRV mode.
PS (cmH
O)
2
Positive airway support pressure value during inspiratory phase.
Sigh. Ampl. (%)
Sigh. Percentage increase of the set Vti.
Sigh. Int. (b)
Sigh. Activation frequency.
Ti (s)
Time that defines the ventilator inspiration duration. The values
can be set based on the set RR.
4-32
User manual, DU322210210