Entering Global System Data - GE DigitalFlow XGS868i Programming Manual

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

1.9.1 Entering Global System Data

This menu enables you to select the system units that the XGS868i will use when channel data is added, subtracted or
averaged together.
1. In the Keypad Program, scroll to PROG and press [Enter].
2. In the PROG menu, scroll to GLOBL and press [Enter].
3. In the Global PROGRAM menu, scroll to SYSTM and press [Enter].
4. At the System Units prompt, scroll to Eng to display measurements in English units or to Metrc for metric units and
press [Enter].
5. Scroll to the desired Pressure Units (absolute or gauge) and press [Enter].
6. Do one of the following:
If absolute was selected, proceed immediately to Step 7.
If gauge was selected, enter the desired Atmospheric Pressure, press [Enter] and proceed to the next step.
7. Do one of the following:
1-Channel meter, go to Procedure Options on page 22.
2-Channel meter, proceed to Selecting Volumetric Units below.
1.9.1a
Selecting Volumetric Units
1. Scroll to the desired Volumetric Units for the flow rate display and press [Enter]. Available units are listed in
Table 5.
2. Scroll to the desired Time units for the flow rate display and press [Enter].
3. Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point in the volumetric flow
rate display) and press [Enter].
DigitalFlow™ XGS868i Programming Manual
Table 5: Available Volumetric/Totalizer Units
English
ACF = Actual Cubic Feet
KACF = Thousands of ACF
MMACF = Millions of ACF
SCF = Standard Cubic Feet
KSCF = Thousands of SCF
MMSCF = Millions of SCF
Metric
ACM = Actual Cubic Meters
KACM = Thousands of ACM
MMACM = Millions of ACM
SCM = Standard Cubic Meters
KSCM = Thousands of SCM
MMSCM = Millions of SCM
Chapter 1. Programming Site Data