C.5.3b Pipe OD
4. Click on the appropriate Pipe OD Unit type in the center pane from the list shown in Table 19. Then enter the
known pipe outside diameter or circumference in the right pane and click [Next Item].
Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the transducer
installation site. The data may also be obtained from standard pipe size tables found in Sound Speeds and Pipe Size
Data (914-004).
inches
feet
circum.in. = pipe circumference in inches
circum.ft = pipe circumference in feet
5. Enter the known Pipe Wall Thickness (in inches or mm) in the right pane and click [Next Item].
C.5.3c
Path and Axial Lengths
6. Click on the appropriate Path Length unit type in the center pane. Then enter the path length of the ultrasonic signal
in the right pane and click [Next Item].
Note: If a spoolpiece was ordered with the meter, the transducer signal path length
(L)
length
are engraved on the flowcell and/or are included in the documentation supplied with the meter. For
on-site transducer installations, refer to Appendix C, Measuring P and L Dimensions, in the Startup Guide for
instructions.
7. Click on the appropriate Axial Length unit type in the center pane. Then enter the axial length of the ultrasonic
signal in the right pane and click [Next Item].
C.5.3d Fluid Type
8. Double-click on the appropriate Fluid Type, either Steam or Other.
If you select Other, PanaView asks for the Fluid Soundspeed. Enter the appropriate soundspeed (in ft/sec) and
click [Next Item].
C.5.3e
Reynolds Correction
9. Double-click on the appropriate choice to indicate whether you want Reynolds Correction.
If Off is selected, enter the Calibration Factor and click on [Next Item].
If On is selected, enter the Kinematic Viscosity and click [Next Item]. Then enter the Calibration Factor and click
[Next Item].
PanaView returns to the
DigitalFlow™ XGS868i Programming Manual
Table 19: Available Pipe OD Units
English
Channel PROGRAM
menu. You have completed entering pipe parameters.
Appendix C. Programming the XGS868i Using PanaView™
Metric
mm
meters
circum.mm = pipe circumference in millimeters
circum.m = pipe circumference in meters
(P)
and the transducer signal axial
101