GE DigitalFlow XGS868i Programming Manual page 42

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

Chapter 1. Programming Site Data
1.9.2g
Totalizer Outputs
The totalizer output issues one pulse per selected volume of flow. The meter produces a pulse each time the
programmed amount of flow passes through the pipe. Refer to Figure 31 on page 80, and complete the following steps
to set up the totalizer outputs of an option card installed in
Accessing the Totalizer Output
1. In the Keypad Program, scroll to PROG and press [Enter].
2. In the PROG menu, scroll to GLOBL and press [Enter].
3. In the Global PROGRAM menu, scroll to I/O and press [Enter].
4. Scroll to OPTN and press [Enter].
5. Scroll to Slot 1 and press [Enter].
6. Scroll to the desired Output and press [Enter].
7. Do one of the following:
If you selected Off, the XGS868i returns to the I/O window.
Go to Procedure Options on the following page.
If you selected TTLZR, and are using a
1-Channel meter, proceed to Step 2 on the next page.
2-Channel meter, proceed to Step 1 on the next page.
Note: Output letters (A, B, etc.) correspond to the output wiring terminal numbers from top to bottom (e.g. Output A
corresponds to pins 1,2 and 3; Output B corresponds to pins 4, 5 and 6; etc.).
32
Slot 1
:
DigitalFlow™ XGS868i Programming Manual