1.8.3 Setting Response Time - V Averaging

Use this option to specify the number of readings that occur before the meter will respond to a step change in flow rate.
In general, the smaller the number of readings, the less steady the display will appear. Complete the following steps to
set the response time:
IMPORTANT:
Do not change this number unless instructed to do so by GE Measurement & Control personnel.
1. In the Keypad Program, scroll to PROG and press [Enter].
2. In the PROG menu, scroll to CH1 or CH2 and press [Enter].
3. In the Channel PROGRAM menu, scroll to Setup and press [Enter].
4. Scroll to AVRG and press [Enter].
5. Scroll to the Response Time (in number of readings) and press [Enter]. For best results, select a value of 30 to
ensure the most stable signal.
Procedure Options
After completing the above steps, the XGS868i returns to the Channel SET UP window. Do one of the following:
To continue programming, refer to Appendix A, Menu Maps, to navigate to the desired menu.
To leave the Keypad Program, press [Escape] four times.
DigitalFlow™ XGS868i Programming Manual
Chapter 1. Programming Site Data
17