GE DigitalFlow XGS868i Programming Manual page 139

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

C.6.3b Requesting Parameters Using MODBUS (cont.)
Notes:
1. Clear Totalizers: flag from the 8051 to clear totalizers in the 68332 memory.
2. Error Value: see table in XMT868i manual for error codes
3. Average:
average of channel 1 and channel 2 if both channels out of error,
channel 1 value if channel 2 is in error,
channel 2 value if channel 1 is in error,
zero if both channels in error.
4. Average Error Status:
0 = both in error
1 = chan 2 in error,
2 = chan 1 in error,
3 = both ok
5. MODBUS baud rate:
5 = 2400, 6 = 4800, 7 = 9600
6. MODBUS parity:
0 = none, 1 = odd, 2 = even
7. MODBUS stop bits:
1 = 1 stop bit, 2 = 2 stop bits
8. General:
Registers are written if corresponding functions are actuated by the user. Registers for unactuated functions are
initialized to zero at startup.
DigitalFlow™ XGS868i Programming Manual
Appendix C. Programming the XGS868i Using PanaView™
129