Chapter 1. Programming Site Data
1.9.2h
Frequency Outputs
The frequency output issues a continuous signal with a frequency proportional to the selected measurement. Refer
tFigure 31 on page 80 and complete the following steps to set up the totalizer outputs of an option card installed in
1
:
Accessing the Frequency Output
1. In the Keypad Program, scroll to PROG and press [Enter].
2. In the PROG menu, scroll to GLOBL and press [Enter].
3. In the Global PROGRAM menu, scroll to I/O and press [Enter].
4. Scroll to OPTN and press [Enter].
5. Scroll to Slot 1 and press [Enter].
6. Scroll to the desired Output and press [Enter].
7. Do one of the following:
If you selected Off, the XGS868i returns to the Input/Output window. Go to Procedure Options on the following
page.
If you selected Frequency, and are using a
1-Channel meter, proceed to Step 2 on the next page.
2-Channel meter, proceed to Step 1 on the next page.
Note: Output letters (A, B, etc.) correspond to the output wiring terminal numbers from top to bottom (e.g. Output A
corresponds to pins 1,2 and 3; Output B corresponds to pins 4, 5 and 6; etc.).