Chapter 1. Programming Site Data
1.7.4 Temperature Input
The Model XGS868 can use either a fixed temperature value or a live temperature input to calculate the steam density
for the mass flow rate display.
3. Scroll to a Fixed temperature value or to set up the option card in Slot 1 that will supply the live temperature input
and press [Enter].
Note: If Slot 1 contains an activated option card with an analog input assigned to
appears as an option at the above prompt. If the process temperature is stable, a fixed value may be used, but
most applications require a live temperature input. If there is no active option card for temperature, the meter
assumes you are using a fixed temperature.
•
If you selected Fixed, proceed to Step 4.
•
If you selected Slot 1, proceed to Step 5.
4. Enter the known Fixed Temp. (process temperature) and press [Enter]. The meter will accept values from -328 to
1,832°F (-200 to 1,000
5. Scroll to Input A or Input B and press [Enter]. The inputs were labeled during setup.
Note: The set up of Input A is used as an example. Identical procedures would be used to set up Input B.
1.7.5 Base Temperature
6. Use the arrow keys to enter the Base Temperature and press [Enter]. The ratio of this value to the actual
temperature is used to calculate the standard mass flow rate.
•
No
If you selected
If you selected Pressure as the Steam Input Type, proceed to Step 7 below.
•
If you selected Temperature as the Steam Input Type, proceed to step 10 on the next page.
•
1.7.6 Pressure Input
7. Scroll to a Fixed pressure value or to set up the option card in Slot 1 that will supply the live pressure input and
press [Enter].
Note: If Slot 1 contains an activated option card with an input assigned to
above prompt. If the process pressure is stable, a fixed value may be used, but most applications require a live
pressure input. If there is no active option card for pressure, the meter assumes you are using a fixed pressure.
If you selected Fixed, proceed to Step 8.
•
If you selected Slot 1, proceed to Step 9.
•
8. Enter the known Fixed process Pressure and press [Enter]. The meter will only accept values from 0 to 5,000 psia.
Proceed to Base Pressure in Step 10.
9. Scroll to Input A or Input B and press [Enter]. The inputs were labeled during setup.
Note: The set up of Input A is used as an example. Identical procedures would be used to set up Input B.
12
o
C). Proceed to Base Temperature in step 6.
at Assume Saturation, proceed to Step 7 below.
Temperature
or an
Pressure, Slot 1
appears as an option at the
DigitalFlow™ XGS868i Programming Manual
RTD
input, Slot 1