Option Card Alarm Relays
Complete the following steps to set up the alarm relays of an option card installed in
page 142):
1. Double-click on the desired Output (A, B, C or D).
Note: The set up of alarm A is used here as an example. Identical procedures would be used to set up the additional
alarms.
2. Double-click on Off to disable Output A and return to step 1, or on High, Low or Fault to select the desired alarm
type.
3. Double-click on No for standard alarm operation, or on Yes for Failsafe operation. (See Chapter 1, Installation, of
the Startup Guide for failsafe wiring instructions.)
For a 1-Channel meter, proceed to step 5.
4. Double-click on the desired Channel option. See Table 28 for a description of the channel options available.
Note: If
FAULT
was selected as the alarm type, the next two prompts do not appear. You have completed entering data
in this option.
5. Double-click on the desired Measurement Name. (Table 27 on page 117 provides a description of the available
parameters.)
Note: The measurement units that appear in these prompts are those selected in the
this section.
6. Enter a value for the Trigger Point of the alarm and click [Next Item].
7. Double-click on the type of Error (flow, non-flow or both) that will trigger the fault alarm and press [Enter].
After responding to the above prompt, the meter returns to step 1. Click [Exit Page] to proceed to another menu or
option.
DigitalFlow™ XGS868i Programming Manual
Table 28: Channel Options
Option
Channel 1
Channel 2
Add Channels
Subtract Channels
Average Channels
Appendix C. Programming the XGS868i Using PanaView™
Slot 1
Description
Channel 1
Channel 2
CH1+CH2
CH1-CH2
(CH1+CH2)/2
Global-System
(refer to Figure 55 on
menu earlier in
121