Appendix C. Programming the XGS868i Using PanaView™
C.5.4f
Pressure Input
1. Double-click on a Fixed pressure value or to set up the option card in Slot 1 that will supply the live pressure input.
Note: If Slot 1 contains an activated option card with an input assigned to
above prompt. If the process pressure is stable, a fixed value may be used, but most applications require a live
pressure input. If there is no active option card for pressure, the meter assumes you are using a fixed pressure.
2. Proceed to one of the following steps:
If you selected Fixed - proceed to Step 3.
If you selected Slot 1 - proceed to Step 4.
3. Enter the known Fixed process Pressure and click [Next Item]. The meter will only accept values from 0 to 5,000
psia. Proceed to Base Pressure below.
4. Double-click on Input A or Input B. The inputs were labeled during setup.
Note: The set up of Input A is used as an example. Identical procedures would be used to set up Input B.
C.5.4g Base Pressure
1. Enter the Base Pressure and click [Next Item]. The ratio of this value to the actual pressure is used to calculate the
standard mass flow rate.
C.5.4h Low Pressure Switch
1. Click on Yes or No to activate or deactivate the Low Pressure Switch software function and click [Next Item].
If you selected Yes, enter the Pressure Limit, the low pressure switch set point, and click [Next Item]. The
acceptable range is 0 to 5,000 psia. The meter will stop taking readings if the pressure drops below this value.
If you selected No, go to Quality Input on page 105.
104
Pressure, Slot 1
appears as an option at the
DigitalFlow™ XGS868i Programming Manual