C.6.1 Entering Global-System Data
While completing these instructions, see the menu map in Figure 54 on page 141 in Appendix D, PanaView Menu
Maps.
1. In the Site Data Menu, double-click on the Global entry.
2. Highlight and double-click on the System option in the center pane. The window now appears similar to Figure 45.
3. Double-click on the desired System Units selection (either metric or English). The XGS868i will display all
parameters and measurements in the designated units.
4. Double-click on the desired Pressure Units (absolute or gauge) and click [Next Item].
a. If gauge was selected, enter the desired Atmospheric Pressure, and click [Next Item].
5. Do one of the following:
For a single-channel XGS868i, the program returns to the
For a 2-channel meter, proceed to step 6 on the next page.
DigitalFlow™ XGS868i Programming Manual
Figure 45: The System Option in the Global Menu
Appendix C. Programming the XGS868i Using PanaView™
menu. Click [Exit Page].
Global PROGRAM
111