Receptacle Plug
Figure 7-12.
Model 3050-AMS Module
reintalation

Replacing the pilot valve manifold

1.
Position the pilot valve manifold on the electronic module mounting
plate. Attach with the four screws you removed earlier.
2.
Attach the tubing you identified and removed earlier. Make sure that
you are attaching the correct tubing to each valve.
3.
Reattach the pilot valve manifold assembly to the sample manifold
plate using the Philips-head captured screws.

Reinstalling the Electronic Module

Once you have serviced the various components in the electronics mod-
ule, reassemble the analyzer.
1.
Carefully reposition the electronic module above and to the front of
the analyzer. Lower the module into the analyzer so the screws at the
four corners can be attached.
2.
Attach the mounting plate of the electronic module to the analyzer as-
sembly using the four captured screws you removed earlier.
Use Philips
screwdriver
Screws (large)
Receptacle Plug
Spring Loaded
Captured Screws
Use slotted
screwdriver
Screws (small)
|
Service & Parts
7-15